Automatic Flash Point Tester

What is the Flash Point?

Flash Point is the lowest temperature, as measured in the prescribed manner, of the test portion corrected to a barometric pressure of 101,3 kPa, at which application of an ignition source causes the vapour of the test portion to ignite momentarily and the flame to propagate across the surface of the liquid under the specified conditions of test.

BGD 240 Automatic Flash Point Tester is designed according to rapid equilibrium closed cup method. It can rapidly test flash point for paints, including water-borne paints, varnishes, binders for paints and varnishes adhesives, solvents, petroleum, and related products.

It conforms to ISO 3679ISO 3680ASTM D 3278ASTM D3828 etc 

Characters

 Can detect the flash point at a specific temperature between -35 ℃ and 105 ℃ (BGD 240/1)or RT  to 300 ℃(BGD 240/2)at average rate heating.

 All functions are achieved through touch screen and digital display.

 Built in two test proceduresrapid balanced mode and average rate heating mode

 Various parameters, such as the heating mode, time, temperature and the average rate heating can be set by the user themselves.

 Planning board action, the decline of the ignition, the detection of flash point is done automatically.

 The gas ignition sources can be provided directly by the laboratory, you can also use the portable butane gas tank.

 Built-in cooling module is used to shorten the cooling time between the two tests.

 Amend barometric pressure automatically 



BGD 240-NEW-02.jpg


BGD 240-NEW-03.jpg

Main Technical Parameters

Ordering Information

Parameter

BGD 240/1 Low Temp.

Automatic Flash Point Tester

BGD 240/2 Normal Temp.

Automatic Flash Point Tester

Measurement Range

-35105

Room Temp.RT)~300

          Tested Flash Point
      Temperature Accuracy

±0.5℃(≤100℃);±2℃(>100℃)

Sample volume

2ml

4ml

Cooling Method

Built-in cooling module (with external cooling   water, flow rate1L/min)

Air-cooled

Heating Rate

User-defined, with a maximum   heating rate of 5°C per minute.

Output Interface

USB, U Disc

Operate Temperature

15℃~35

Power Supply

220/240V50/60Hz

Overall Size

460mm×330mm×350mmL×W×H

Net Weight

15 Kg

Come with Accessories

Steam source regulation componentSilicone tubeinjector

                                      ThermometerO-Type Seal Ring5pcs

Optional Accessories

Injector for high viscosity sampleStand oil of Flash   Point


Introduction

Subject:
Company:*
Name:
E-mail:*
Country:*
Comments:
Verification Code:Click ChangeCan't see clearly. Change it?
 
Send message
Contact USMore

Biuged Precise Instruments (Guangzhou) Co.,Ltd

AddNO.15, Fuzhong Rd., Zhongxin Town, 

Zengcheng District, Guangzhou.China

Tel+86-20-32955999;+86-20-32955820

Fax+86-20-32955819 

E-mailjack@biuged.com

Website http://www.biuged.com 


Biuged Instruments Co.,Ltd.

Add:RM.310, NO.1, Sicheng Rd.,
Tianhe District, Guangzhou.China

Tel:+86-20-82169666

Fax:+86-20-32955818

E-mail:jack@biuged.com





Biuged Precise Instruments (Shanghai) Co.,Ltd

Add:Room 906, 9th Floor, Zhongcai Industrial Park, No.1111 

Zhu Feng Road, Qingpu District, Shanghai, China

Tel:+86-21-59240298

Fax:+86-21-59249719

E-mail:xjx@biuged.com

  Biuged Laboratory Instruments (Guangzhou) Co., Ltd.   Guangdong ICP record number:11044672